Replacing a door seal on a steam steriliser or autoclave is an important part of maintaining the equipment's sealing performance. Correct installation is essential to ensure the seal sits evenly within the door channel and provides an effective seal when the door is closed.
At TYM Seals & Gaskets, we manufacture and supply replacement door seals for a wide range of steam sterilisers and washer-disinfectors. To help engineers and maintenance teams with the fitting process, we've put together this step-by-step guide explaining how to install a replacement steriliser door seal correctly.
Before You Start: Important Installation Advice
Before fitting your replacement door seal, there are a few important points to consider:
-
Check the seal orientation: Ensure the joint is positioned away from the gasket charge port, which is normally located at the top of the door channel.
-
Avoid twisting: Keep the seal correctly aligned as you work around the door opening. Twisting can affect how it sits within the channel.
-
Do not pull or stretch the seal: The seal should be compressed into position rather than pulled to make it fit.
-
Position the joint correctly: The joint should sit approximately 2 inches to the left or right of the top-centre position of the door opening.
Following these recommendations will help ensure the seal is evenly distributed around the channel.

How to Install a Steam Steriliser Door Seal
The following five steps explain how to fit a replacement door seal into a steriliser door channel.
Step 1 – Position the Starting Point
Begin by selecting your starting position (Point 1 on the installation diagram). Push approximately 2 inches of the seal into the door channel, ensuring it is sitting correctly without being twisted or stretched.
Step 2 – Locate the Opposite Position
Move to the opposite side of the door opening and identify the centre point of the remaining loose seal. Insert approximately 2 inches into the channel at Point 2.
This helps establish the correct positioning of the seal around the opening before continuing.
Step 3 – Secure the Intermediate Positions
Next, locate the halfway positions between Points 1 and 2. Insert approximately 2 inches of the seal into the channel at Points 3 and 4.
At this stage, the seal should be held in position at four evenly distributed points around the door opening.
Step 4 – Add Further Fixing Points
Repeat the process by identifying the remaining halfway positions around the seal.
Insert approximately 2 inches of the seal into the channel at each of the four additional positions (Points 5, 6, 7 and 8).
The seal should now be held at eight separate points, helping to maintain an even distribution before completing the installation.
Step 5 – Complete the Installation
With the seal secured at all eight positions, work around the remaining sections of the door channel, carefully pressing the seal into place.
Use a firm hand movement to compress the seal into the channel rather than pulling or stretching it.
Important: Replacement seals may be manufactured slightly larger than the door opening to allow for the correct fit. This is intentional and helps ensure the seal sits properly once installed. Avoid removing the excess length by stretching or forcing the seal into position.
Once fully installed, check that the seal is seated evenly around the entire channel and that the joint is correctly positioned.
Always follow the steriliser manufacturer's installation and commissioning procedures before returning the equipment to service.
Replacement Steam Steriliser Door Seals from TYM
TYM Seals & Gaskets manufactures and supplies replacement door seals for a wide range of steam sterilisers and autoclaves, supporting hospitals, healthcare facilities, service engineers and equipment maintenance companies.
We offer sealing solutions for many leading steriliser manufacturers, providing a reliable alternative to branded replacement parts at competitive prices.
Whether you require a standard replacement seal or need assistance identifying the correct size and profile for your equipment, our experienced team can help.
Contact TYM Seals & Gaskets to discuss your requirements or enquire about our range of steam steriliser door seals.